QA of dynamic MLC based on EPID portal dosimetry.
2010
Abstract Purpose Dynamic delivery of intensity modulated beams (dIMRT) requires not only accurate verification of leaf positioning but also a control on the speed of motion. The latter is a parameter that has a major impact on the dose delivered to the patient. Time consumed in quality assurance (QA) procedures is an issue of relevance in any radiotherapy department. Electronic portal imaging dosimetry (EPID) can be very efficient for routine tests. The purpose of this work is to investigate the ability of our EPID for detecting small errors in leaf positioning, and to present our daily QA procedures for dIMRT based on EPID. Methods and materials A Varian 2100 CD Clinac equipped with an 80 …